FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2182746
·
Received July 8, 2011
Report
- Report Number
- 1720753-2011-08843
- Event Type
- Malfunction
- Date Received
- July 8, 2011
- Date of Event
- June 17, 2011
- Report Date
- July 8, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE ORBITAL AND ROTATION IRIS POTENTIOMETERS AND THE SERVO'S WERE REPLACED AND CALIBRATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED MOTION ERROR MESSAGES. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |