FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 2182736 · Received July 28, 2011

Report

Report Number
2050012-2011-03865
Event Type
Malfunction
Date Received
July 28, 2011
Date of Event
July 1, 2011
Report Date
July 5, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K060256
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A BEC FIELD SERVICE ENGINEER (FSE) REMOVED AND DISASSEMBLED THE GRAVITY SUMP. FSE DID NOT NOTE ANY ISSUES WITH O-RING, FLOAT AND VALVE. FSE REPLACED THE FLOAT SWITCH AND PRIMED THOROUGHLY. SUMP IS DRAINING NORMALLY. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) TO REPORT GRAVITY SUMP IS INTERMITTENTLY OVERFLOWING. NO INJURY OR EXPOSURE TO FLUID WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER FOR CLINICAL USE. JJE BECKMAN COULTER INC. DXC 600I N/A

Patients

Seq Age Sex Outcome Treatment
1