FDA Adverse Event Injury Summary report: N

DOUBLE MOBILITY LINER

MDR report key: 21825184 · Received April 11, 2025

Report

Report Number
3005180920-2025-00292
Event Type
Injury
Date Received
April 11, 2025
Date of Event
March 24, 2025
Report Date
April 11, 2025
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
MEH
PMA / PMN Number
K092265
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 26 MARCH 2025: LOT 2347552: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 20/02/2024. EXPIRATION DATE: 05/02/2029. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ADDITIONAL IMPLANT REVISED: BATCH REVIEW PERFORMED ON 26 MARCH 2025 ON BALL HEADS: MECTACER 01.29.202 BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 28 SIZE M 0 (K112115) LOT. 2414769 LOT 2414769: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 10/06/2024. EXPIRATION DATE: 23/05/2029. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. THERE IS NO INDICATION THAT ANY POTENTIAL ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AND THE DEVICE HISTORY RECORD REVIEW DOES NOT INDICATE ANY POTENTIAL MANUFACTURING RELATED ISSUE.

Description of Event or Problem · 0

THE PATIENT CAME IN REPORTING PAIN. REVISION SURGERY PERFORMED DUE TO DISSOCIATION OF THE HEAD FROM THE DOUBLE MOBILITY LINER (INTRA-PROSTHETIC DISLOCATION). ABOUT 6 MONTHS AFTER THE PRIMARY SURGERY, THE SURGEON REVISED THE MEDACTA CUP AND LINER TO COMPETITOR COMPONENTS AND REVISED THE MEDACTA HEAD WITH A MEDACTA HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY. THE SURGEON HAD TO REVISE THE CUP BECAUSE A SEMI CONSTRAINED LINER IS NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1104175 DOUBLE MOBILITY LINER LINER: VERSAFITCUP DM DOUBLE MOBILITY HC LINER Ø 48/28 MEH MEDACTA INTERNATIONAL SA 01.26.2848MHC 2347552

Patients

Seq Age Sex Outcome Treatment
1 57 YR Female Required Intervention