COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-15253
- Event Type
- Malfunction
- Date Received
- July 28, 2011
- Date of Event
- July 5, 2011
- Report Date
- July 5, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A FAILURE CODE OF 310:422:558:000 WAS CONFIRMED AND DUPLICATED DURING PRODUCT EVALUATION. THE ROOT CAUSE WAS ASSIGNED TO THE USER INTERFACE PRINTED CIRCUIT BOARD. THE USER INTERFACE PRINTED CIRCUIT BOARD WAS REPLACED IN ORDER TO CORRECT THIS CONDITION. THIS INVOLVED A COLLEAGUE P1.5 INFUSION PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.92.
THE DEVICE IS REPORTED TO BE AVAILABLE FOR EVALUATION. UPON COMPLETION OF BAXTER'S INVESTIGATION A FOLLOW UP MEDWATCH WILL BE SUBMITTED (B)(4).
THIS IS A REPORT OF A COLLEAGUE INFUSION PUMP WITH A FAILURE CODE 310:422:558:000. IT IS UNKNOWN WHEN THE REPORTED CONDITION OCCURRED. PATIENT INVOLVEMENT WAS NOT REPORTED. NO PATIENT INJURY OR MEDICAL INTERVENTION OCCURRED. THE SOFTWARE VERSION FOR THE DEVICE IS CURRENTLY NOT KNOWN. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |