FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2182472 · Received July 28, 2011

Report

Report Number
6000001-2011-15253
Event Type
Malfunction
Date Received
July 28, 2011
Date of Event
July 5, 2011
Report Date
July 5, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A FAILURE CODE OF 310:422:558:000 WAS CONFIRMED AND DUPLICATED DURING PRODUCT EVALUATION. THE ROOT CAUSE WAS ASSIGNED TO THE USER INTERFACE PRINTED CIRCUIT BOARD. THE USER INTERFACE PRINTED CIRCUIT BOARD WAS REPLACED IN ORDER TO CORRECT THIS CONDITION. THIS INVOLVED A COLLEAGUE P1.5 INFUSION PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.92.

Additional Manufacturer Narrative · 1

THE DEVICE IS REPORTED TO BE AVAILABLE FOR EVALUATION. UPON COMPLETION OF BAXTER'S INVESTIGATION A FOLLOW UP MEDWATCH WILL BE SUBMITTED (B)(4).

Description of Event or Problem · 1

THIS IS A REPORT OF A COLLEAGUE INFUSION PUMP WITH A FAILURE CODE 310:422:558:000. IT IS UNKNOWN WHEN THE REPORTED CONDITION OCCURRED. PATIENT INVOLVEMENT WAS NOT REPORTED. NO PATIENT INJURY OR MEDICAL INTERVENTION OCCURRED. THE SOFTWARE VERSION FOR THE DEVICE IS CURRENTLY NOT KNOWN. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1