FDA Adverse Event Malfunction Summary report: N

INFUSOMAT®

MDR report key: 21824581 · Received April 11, 2025

Report

Report Number
9610825-2025-00211
Event Type
Malfunction
Date Received
April 11, 2025
Date of Event
March 3, 2025
Report Date
May 1, 2025
Manufacturer
B BRAUN MELSUNGEN AG
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN INTERNAL REPORT NUMBER (B)(4). A FOLLOW-UP REPORT WILL BE PROVIDED AFTER THE EXAMINATION RESULTS ARE AVAILABLE. PLEASE NOTE, THIS DEVICE IS DISTRIBUTED OUTSIDE THE US AND NO GUIDED INFORMATION EXISTS. IT IS BEING REPORTED DUE TO A SIMILAR DEVICE BEING MARKETED WITHIN THE US WITH THE FOLLOWING INFO: (B)(4). PREMARKET SUBMISSION # K083689, K142596, K191910.

Additional Manufacturer Narrative · 0

THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN INTERNAL REPORT NUMBER (B)(4). PLEASE NOTE, THIS DEVICE IS DISTRIBUTED OUTSIDE THE US AND NO GUDID INFORMATION EXISTS. IT IS BEING REPORTED DUE TO A SIMILAR DEVICE BEING MARKETED WITHIN THE US WITH THE FOLLOWING INFO: UDI NUMBER: (B)(4). PREMARKET SUBMISSION # K083689, K142596, AND K191910. THE DEVICE WAS NOT AVAILABLE. ONLY THE HISTORY DATA WAS SENT FOR INVESTIGATION. RESULTS: THE DEVICE HISTORY FILES WERE ANALYZED. A SPACE LINE WAS SELECTED, AND THE INFUSION STARTED WITH A RATE OF 26ML/H AND A VOLUME OF 625ML OVER 24 HOURS. AFTER 24 HOURS WAS REACHED AND THE KVO-MODE (KEEP VEIN OPEN) STARTED. THE INFUSION WAS STOPPED, AND THE LINE WAS EXTRACTED. NO OTHER ABNORMALITIES WERE FOUND. JUDGMENT: THE COMPLAINT COULD NOT BE CONFIRMED.

Description of Event or Problem · 0

ACCORDING TO THE EVENT DESCRIPTION: THE INFUSION OVER RUN. IT WAS SUPPOSED TO BE INFUSED OVER 24 HOURS. AND IT IS DUE FOR COMPLETION ON (B)(6) 2025 AT 1700 HOURS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1092962 INFUSOMAT® PUMP, INFUSION FRN B BRAUN MELSUNGEN AG 8713050 96974551G8

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown