OLYMPIC COOL-CAP SYSTEM
Report
- Report Number
- 3018859-2011-00002
- Event Type
- Malfunction
- Date Received
- July 11, 2011
- Date of Event
- June 13, 2011
- Report Date
- July 11, 2011
- Manufacturer
- NATUS MEDICAL INC.,
- Product Code
- MXM
- PMA / PMN Number
- P040025
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NURSE
Narratives
COOL-CAP SYSTEM IS BEING RETURNED TO NATUS MEDICAL INC FOR TESTING. A F/U REPORT WILL BE SUBMITTED TO THE FDA WHEN THE DEVICE INVESTIGATION / TESTING IS COMPLETED. ADDITIONAL SERIAL #: (B)(4).
DURING TREATMENT OF AN INFANT, AN OLYMPIC COOL-CAP SYSTEM EXHIBITED A SHUTDOWN CHARACTERIZED BY A FROZEN DISPLAY SCREEN. NO AUDIBLE OR VISIBLE ALARMS WERE OBSERVED. DEVICE FIRST EXPERIENCED THIS EVENT AT APPROX 22 HRS INTO THE 76-HR TREATMENT. SYSTEM WAS FROZEN FOR 45 MINUTES BEFORE IT WAS DISCOVERED. RECTAL TEMP WAS 33.7C WHEN ISSUE WAS DISCOVERED (WITHIN 0.3C OF THE TARGET 34.0-35.0C RANGE). DEVICE WAS REBOOTED AND TREATMENT CONTINUED. DEVICE EXHIBITED A SECOND FROZEN SCREEN; TREATMENT CONTINUED AFTER REBOOTING THE DEVICE. AFTER EXHIBITING A THIRD FROZEN SCREEN, THE PT WAS SHIFTED TO A SECOND COOL-CAP DEVICE. THE REMAINDER OF THE 3-DAY TREATMENT WAS COMPLETED WITHOUT FURTHER INCIDENT. THERE WAS NO PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OLYMPIC COOL-CAP SYSTEM | COOLING CAP (INFANTS) | MXM | NATUS MEDICAL INC., | 60010 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |