FDA Adverse Event Malfunction Summary report: N

VOYAGER NC CORONARY DILATATION CATHETER

MDR report key: 2182406 · Received July 28, 2011

Report

Report Number
2024168-2011-05345
Event Type
Malfunction
Date Received
July 28, 2011
Date of Event
July 4, 2011
Report Date
July 4, 2011
Manufacturer
AV-TEMECULA-CT
Product Code
LOX
PMA / PMN Number
P810046
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. BALLOON MATERIAL RUPTURES CAN BE AFFECTED BY NUMEROUS FACTORS INCLUDING, BUT NOT LIMITED TO, BALLOON DAMAGE DURING MANUFACTURING, MATERIALS, HANDLING, LESION CALCIFICATION AND TORTUOSITY, AN INTERACTION WITH A PREVIOUSLY IMPLANTED STENT, INSUFFICIENT PREPARATION PRIOR TO USE OR FROM INTERACTIONS WITH OTHER DEVICES. IN THIS CASE, THE LESION WAS CHARACTERIZED AS MILDLY TORTUOUS, MILDLY CALCIFIED AND 99% STENOSED, WHICH MAY HAVE CONTRIBUTED TO THE REPORTED DIFFICULTY. IT IS POSSIBLE THAT THE BALLOON MATERIAL BECAME DAMAGED (SCRATCHED) FROM INTERACTIONS WITH OTHER DEVICES AND/OR THE TORTUOUS AND CALCIFIED LESION, SUCH THAT THE BALLOON RUPTURED UPON INFLATION. RETURN OF THE VOYAGER NC MAY HAVE ULTIMATELY AIDED THE INVESTIGATION IN DETERMINING A CAUSE FOR THE EXPERIENCED RUPTURE. TO ENSURE THIS TYPE OF DAMAGE IS NOT A RESULT OF A POTENTIAL PRODUCT RELATED DEFICIENCY, ALL DILATATION CATHETERS ARE 100% VISUALLY INSPECTED AND LEAK TESTED DURING THE MANUFACTURING PROCESS. A SAMPLING OF UNITS IS ALSO DESTRUCTIVELY TESTED TO VERIFY RATED BURST PRESSURE AND BALLOON INTEGRITY. IN THIS CASE, ALTHOUGH THERE WAS NO REPORT OF ANY LEAKS NOTED DURING DEVICE PREPARATION, ADDITIONAL INFORMATION RECEIVED STATED THAT THE DEVICE WAS PREPPED INSIDE THE BODY. IT SHOULD BE NOTED THE VOYAGER NC INSTRUCTIONS FOR USE CAUTIONS: ALL AIR MUST BE REMOVED FROM THE BALLOON AND DISPLACED WITH CONTRAST PRIOR TO INSERTING INTO THE BODY; OTHERWISE, COMPLICATIONS MAY OCCUR. A REVIEW OF THE FINISHED PRODUCT LOT HISTORY RECORD DID NOT REVEAL ANY NONCONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT AND ALL LOT RELEASE TESTING MET MANUFACTURING CRITERIA. FURTHERMORE, A QUERY OF THE COMPLAINT-HANDLING DATABASE DID NOT REVEAL ANY OTHER INCIDENTS REPORTED FOR A BALLOON RUPTURE FROM THIS LOT, SUGGESTING THAT THERE ARE NO LOT SPECIFIC PRODUCT QUALITY DEFICIENCIES. BASED ON THE INFORMATION RECEIVED WITH THIS INCIDENT AND WITHOUT THE PRODUCT TO EXAMINE, A DEFINITIVE CAUSE FOR THE REPORTED BALLOON RUPTURE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LESION WAS IN THE DISTAL RIGHT CORONARY ARTERY, WITH MILD TORTUOSITY, MILD CALCIFICATION, AND A 99% STENOSIS. THE 2.75 X 15 MM VOYAGER NC BALLOON CATHETER WAS ADVANCED FOR PRE-DILATATION WITH NO RESISTANCE FELT; HOWEVER, THE BALLOON RUPTURED AT 6 ATMOSPHERES WHEN INFLATED FOR THE FIRST TIME. THE 2.75 X 15 MM VOYAGER NC BALLOON CATHETER WAS REPLACED WITH A 2.75 X 8 MM VOYAGER NC BALLOON CATHETER AND A XIENCE V STENT WAS IMPLANTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VOYAGER NC CORONARY DILATATION CATHETER CORONARY DILATATION CATHETER LOX AV-TEMECULA-CT 1021461

Patients

Seq Age Sex Outcome Treatment
1 GUIDE WIRE: RUNTHROUGH NSGUIDE CATH: IVUS