FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2182341 · Received July 28, 2011

Report

Report Number
2531779-2011-05416
Event Type
Malfunction
Date Received
July 28, 2011
Date of Event
July 1, 2011
Report Date
July 1, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/27/2011 WITH THE FOLLOWING FINDINGS: THE BOLUS BUTTON WAS OBSERVED TO BE MISSING AND MOISTURE WAS OBSERVED UNDER THE DISPLAY LENS. THE ALL BUTTONS WERE FOUND TO BE UNRESPONSIVE TO BUTTON PRESSES. THE PUMP WAS FOUND TO BE LEAKING FROM THE BOLUS BUTTON. THE PUMP WAS OPENED AND MOISTURE WAS FOUND ON THE PCB AND SCREEN. (B)(6).

Additional Manufacturer Narrative · 1

ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

A FAMILY MEMBER REPORTED THAT AFTER THE PATIENT WENT SWIMMING THE PUMP SCREEN WENT BLANK AND THE KEYPAD STOPPED RESPONDING. THE PATIENT REPORTEDLY WORE THE PUMP ON THE OUTSIDE OF CLOTHING AND DID NOT CLEAN THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 16 YR