FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 21822974 · Received April 11, 2025

Report

Report Number
2955842-2025-14449
Event Type
Malfunction
Date Received
April 11, 2025
Date of Event
March 21, 2025
Report Date
March 21, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE ROOT CAUSE IS ATTRIBUTED TO THE USER, AS DVI WAS NOT PROPERLY CONNECTED. THE CABLES INTENDED TO BE USED WERE NOT THE SPECIFIED.

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) ADDRESSED THE REPORTED EVENT WITH PHONE SUPPORT. THE ISSUE WAS RESOLVED BY THE OPERATING ROOM (OR) STAFF CONTACTING THE COORDINATOR AND GUIDING TO FIND CORRECT CABLES TO RESTORE STRYKER TOWER CONNECTION. NO SITE VISIT WAS CONDUCTED. THE SYSTEM WAS WORKING PROPERLY AND NO ADDITIONAL ACTION WAS REQUIRED AS THE ISSUE WAS RESOLVED.

Description of Event or Problem · 0

REFER TO H11 FOR FOLLOW-UP INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO THE START OF A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE CUSTOMER INFORMED THE TECHNICAL SUPPORT ENGINEER (TSE) THE VIDEO OUTPUT TO STRYKER BOOM STARE WASN¿T WORKING. THE SURGEON CONVERTED TO LAP BEFORE CALLING. THE CUSTOMER STATED THE SYSTEM WAS MOVED OUT OF THE ROOM SO FLOORS COULD BE CLEANED. WHEN THE SYSTEM WAS MOVED BACK IN, THE ONLY DIGITAL VIDEO INTERFACE CABLE THEY COULD FIND DID NOT CONNECT ONTO THE PORT IN VISION SIDE CART. THE PROCEDURE WAS CONVERTED TO LAPAROSCOPIC SURGERY WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: FOLLOW-UP ATTEMPT #: ON 3/27/2025, THE DAVINCI COORDINATOR (B)(6) DOES NOT HAVE ANY INFORMATION OR FURTHER DETAILS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2350165 DAVINCI XI VISION SIDE SYSTEM NAY INTUITIVE SURGICAL, INC 381121-45 N/A

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES