VASOVIEW HEMOPRO 2
Report
- Report Number
- 2242352-2025-0000443
- Event Type
- Malfunction
- Date Received
- April 10, 2025
- Date of Event
- March 11, 2025
- Report Date
- April 16, 2025
- Manufacturer
- MAQUET CARDIOVASCULAR LLC
- Product Code
- GEI
- PMA / PMN Number
- K101274
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
TW (B)(4). THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION ON 03/21/2025. AN INVESTIGATION WAS CONDUCTED ON 04/02/2025. A VISUAL INSPECTION WAS CONDUCTED. SIGNS OF CLINICAL USE AND EVIDENCE OF BLOOD WAS OBSERVED. THERE WERE NO VISUAL DEFECTS OBSERVED ON THE INTACT BTT. A 25 CC SYRINGE, FILLED WITH AIR WAS ATTACHED TO THE INFLATION PORT LINE ON THE BTT AND SQUEEZED THE SYRINGE TO INJECT AIR. THE BTT WAS ABLE TO BE INFLATED. A 25CC SYRINGE, FILLED WITH SALINE WAS ATTACHED TO THE CO2 LINE ON THE BTT AND SQUEEZED THE SYRINGE TO SPRAY THE SALINE WITH NO ISSUES. A 25CC SYRINGE, FILLED WITH SALINE WAS ATTACHED TO THE CO2 LINE ON THE BTT AND SQUEEZED THE SYRINGE TO SPRAY THE SALINE. THERE WAS NO BACKFLOW OBSERVED IN THE CO2 LINE. BASED ON THE RETURNED CONDITION OF THE DEVICE AS WELL AS THE EVALUATION RESULTS, THE REPORTED FAILURE "IMPROPER FLOW OR INFUSION" WAS NOT CONFIRMED. THE LOT # 3000458376 HISTORY RECORD REVIEW WAS COMPLETED. THERE WERE NO NCMRS, REWORK, OR DEVIATIONS DOCUMENTED FOR THE REPORTED LOT NUMBER. BASED ON THE DHR/LHR REVIEW RESULTS, IT WAS DETERMINED THAT THERE IS NO RELATION BETWEEN THE BATCH MANUFACTURING PROCESS AND THE REPORTED FAILURE.
(B)(4). THE DEVICE HAS BEEN RETURNED TO THE FACTORY AND IS BEING EVALUATED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETED.
N/A.
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE USING VASOVIEW HEMOPRO 2, DURING DISSECTION THE CUSTOMER GETTING A INSUFFLATION OCCLUSION MESSAGE ON THEIR INSUFFLATION DEVICE. ALL CONNECTIONS CHECKED WITHOUT IMPROVEMENT. THE CUSTOMER SWITCHED OUT THE ENTIRE EVH KIT AND THE PROBLEM WAS RESOLVED. NOR HARM TO PATIENT. NO. PROCEDURE DELAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2340450 | VASOVIEW HEMOPRO 2 | ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES | GEI | MAQUET CARDIOVASCULAR LLC | VH-4000 | 3000458376 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Female |