FDA Adverse Event Malfunction Summary report: N

ACCESS

MDR report key: 2182153 · Received July 28, 2011

Report

Report Number
6000001-2011-15223
Event Type
Malfunction
Date Received
July 28, 2011
Date of Event
April 28, 2011
Report Date
July 1, 2011
Manufacturer
BAXTER HEALTHCARE
Product Code
FPA
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CORRECTION: DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER PERSONNEL, A COLLEAGUE INFUSION PUMP EXPERIENCED AN UPSTREAM OCCLUSION ALARM AFTER ONE MINUTE OR MORE OF DELIVERY PRIOR TO AN OPEN KEY PRESS. THE OPEN KEY PRESS INDICATED THAT AN UNKNOWN SET WAS REPLACED. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THIS COMPLAINT WILL ADDRESS THE POSSIBILITY THAT THE UNKNOWN SET WAS DEFECTIVE CAUSING THE UPSTREAM OCCLUSION ALARM. PATIENT INVOLVEMENT IS UNKNOWN. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED FOR THIS REPORT. THIS WAS NOT REPORTED BY THE CUSTOMER. NO ADDITIONAL INFORMATION IS AVAILABLE. DEVICE EVALUATION: THE SAMPLE WAS NOT AVAILABLE FOR EVALUATION; THEREFORE, THE CONDITION CANNOT BE CONFIRMED OR DUPLICATED AND THE ASSIGNABLE ROOT CAUSE COULD NOT BE DETERMINED. TE DEVICE USED IN THIS SITUATION IS UNKNOWN; THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. IF ADDITIONAL INFORMATION OR SAMPLES BECOME AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Description of Event or Problem · 1

DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER PERSONNEL, A COLLEAGUE INFUSION PUMP WAS FOUND TO HAVE EXPERIENCED A MANUAL TUBE RELEASE OPEN KEY PRESS ALARM, WHICH INTERRUPTED DELIVERY. THIS ALARM MAY HAVE BEEN A FALSE ALARM. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.90. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACCESS SET, ADMINISTRATION, INTRAVASCULAR FPA BAXTER HEALTHCARE

Patients

Seq Age Sex Outcome Treatment
1