FDA Adverse Event Malfunction Summary report: N

IOWA METH DES MOINES IA 1

MDR report key: 2182104 · Received June 23, 2011

Report

Report Number
1718850-2011-00063
Event Type
Malfunction
Date Received
June 23, 2011
Date of Event
May 12, 2011
Report Date
May 26, 2011
Manufacturer
SORIN GROUP USA, INC.
Product Code
DTZ
PMA / PMN Number
PREAMENDMENT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SORIN GROUP USA RECEIVED A REPORT THAT A CRACKED CONNECTOR IN THE CARDIOPLEGIA LINE CAUSED BLOOD TO LEAK AND AIR TO BE ASPIRATED. THIS RESULTED IN A DELAY OF ABOUT 10 MINUTES WHILE THE PT WAS ON PUMP, BUT BEFORE CROSS-CLAMP. THE CONNECTOR WAS REPLACED AND THE CASE CONTINUED WITHOUT FURTHER ISSUES. NO AIR REACHED THE PT AND THERE WERE NO PT COMPLICATIONS DUE TO THE INCIDENT. NO MEDICAL INTERVENTION WAS REQUIRED. THE CONNECTOR WHICH LEAKED WAS RETURNED TO SORIN GROUP USA FOR EVALUATION. WHILE NO CRACK WAS FOUND IN THE CONNECTOR, LEAK TESTING CONFIRMED THE PRESENCE OF A LEAK. A CAPA HAS BEEN INITIATED AND THE ROOT CAUSE DETERMINED. THE CONNECTOR DIMENSIONS WERE OUT OF SPECIFICATION. CORRECTIVE ACTION HAS BEEN IMPLEMENTED BY THE SUPPLIER. THE MFG PROCESS HAS BEEN ADJUSTED TO BRING THE CONNECTOR DIMENSIONS BACK INTO SPECIFICATION. NO FURTHER ACTION IS NECESSARY.

Description of Event or Problem · 1

SORIN GROUP USA RECEIVED A REPORT THAT A CRACKED CONNECTOR IN THE CARDIOPLEGIA LINE CAUSED BLOOD TO LEAK AND AIR TO BE ASPIRATED. THIS RESULTED IN A DELAY OF ABOUT 10 MINUTES WHILE THE PT WAS ON PUMP, BUT BEFORE CROSS-CLAMP. THE CONNECTOR WAS REPLACED AND THE CASE CONTINUED WITHOUT FURTHER ISSUES. NO AIR REACHED THE PT AND THERE WERE NO PT COMPLICATIONS DUE TO THE INCIDENT. NO MEDICAL INTERVENTION WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 IOWA METH DES MOINES IA 1 CUSTOM PERFUSION PACK DTZ SORIN GROUP USA, INC. NA 1033500001

Patients

Seq Age Sex Outcome Treatment
1 80 YR