FDA Adverse Event
Injury
Summary report: N
CAPSURE EPI
MDR report key: 2181957
·
Received July 28, 2011
Report
- Report Number
- 2182208-2011-01073
- Event Type
- Injury
- Date Received
- July 28, 2011
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- DTB
- PMA / PMN Number
- P950024
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT RIGHT VENTRICULAR LEAD HAD HIGH IMPEDANCE MEASUREMENTS OF 7431 OHMS. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSURE EPI | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC, INC. | 4965 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 8 YR | Hospitalization| R | (B)(4) IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACING LEAD| 8042 IMPLANTABLE PULSE GENERATOR |