ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2011-05385
- Event Type
- Malfunction
- Date Received
- July 28, 2011
- Date of Event
- June 28, 2011
- Report Date
- June 28, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
THERE IS NO PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. THE KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE DURING TESTING. DURING INVESTIGATION, MISALIGNED BUTTON CONTACTS WERE OBSERVED. (B)(6).
IT WAS REPORTED THAT ALL KEYPAD BUTTONS (UP ARROW, DOWN ARROW, OKAY, AND CONTRAST) WERE UNRESPONSIVE. THERE IS NO ADDITIONAL INFORMATION AVAILABLE ABOUT THIS COMPLAINT. THE COMPLAINT IS BEING REPORTED BECAUSE UNRESPONSIVE BUTTONS HAVE BEEN KNOWN ON OCCASION TO CONTRIBUTE TO A PATIENT EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |