FDA Adverse Event Malfunction Summary report: N

NC QUANTUM APEX PTCA DILATATION CATHETER

MDR report key: 2181827 · Received July 28, 2011

Report

Report Number
2134265-2011-03038
Event Type
Malfunction
Date Received
July 28, 2011
Date of Event
July 4, 2011
Report Date
July 4, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MANUFACTURER: THE NC QUANTUM APEX BALLOON CATHETER WAS RECEIVED FOR ANALYSIS. THERE WAS BLOOD AND THICK CONTRAST IN THE INFLATION LUMEN AND BALLOON. A PINHOLE WAS FOUND IN THE BALLOON WALL LOCATED ON THE DISTAL END OF THE BALLOON 5MM FROM THE DISTAL TIP. INSPECTION OF THE BALLOON PRESENTED NO IRREGULARITIES IN THE BALLOON MATERIAL OR THE RO MARKERS. THERE IS NO EVIDENCE OF ANY MATERIAL OR MANUFACTURING DEFICIENCIES THAT COULD HAVE CONTRIBUTED TO THIS COMPLAINT. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: (B)(6). (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI) PROCEDURE, A BALLOON RUPTURE OCCURRED. THE 90% STENOSED LESION WAS LOCATED IN THE SEVERELY CALCIFIED AND SEVERELY TORTUOUS RIGHT CORONARY ARTERY. THE PHYSICIAN ADVANCED THE 15MM X 3.50MM NC QUANTUM APEX MR BALLOON TO THE LESION FOR POST DILATION AND UPON INFLATION, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS IS RECORDED AS GOOD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI) PROCEDURE, A BALLOON RUPTURE OCCURRED. THE 90% STENOSED LESION WAS LOCATED IN THE SEVERELY CALCIFIED AND SEVERELY TORTUOUS RIGHT CORONARY ARTERY. THE PHYSICIAN ADVANCED THE 15MM X 3.50MM NC QUANTUM APEX MR BALLOON TO THE LESION FOR POST DILATION AND UPON INFLATION, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS IS RECORDED AS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NC QUANTUM APEX PTCA DILATATION CATHETER CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493912415350 14064539

Patients

Seq Age Sex Outcome Treatment
1