FATHOM GUIDEWIRE
Report
- Report Number
- 2939204-2011-00390
- Event Type
- Malfunction
- Date Received
- July 28, 2011
- Report Date
- July 19, 2011
- Manufacturer
- BOSTON SCIENTIFIC - SALT LAKE CITY
- Product Code
- DQX
- PMA / PMN Number
- K050964
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SE
- Reporter Occupation
- PHYSICIAN
Narratives
EVENTTHE DEVICE HISTORY RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS UPON ITS RELEASE. VISUAL INSPECTION OF THE RETURNED DEVICE DISCOVERED THAT THE GUIDEWIRE HAD SEPARATED AT ITS DISTAL SECTION 178.9CM FROM ITS PROXIMAL END. THE CORE WIRE OF THE GUIDEWIRE, NITINOL (NITI) TUBINGS AND PLATINUM COIL WERE SEPARATED. THE GUIDEWIRE PLATINUM COIL WAS SEVERELY STRETCHED. THE NITI WAS BENT AND SLIGHTLY STRETCHED EXPOSING THE DISTAL END OF THE CORE WIRE. THE GUIDEWIRE WAS BENT IN SEVERAL PLACES. IT WAS ALSO BENT JUST PROXIMAL TO THE SEPARATED SITE. THE GUIDEWIRE'S SEPARATED DISTAL SECTION WAS NOT RETURNED. NO OTHER ANOMALIES WERE NOTED. IT APPEARED THAT THE GUIDEWIRE MIGHT HAVE SEPARATED DURING THE TIP SHAPING PROCESS DUE TO EXCESSIVE FORCE. THE DEVICE DIRECTIONS FOR USE INSTRUCT USERS TO ¿CAREFULLY SHAPE THE GUIDEWIRE TIP WHILE THE OBSERVED STRETCHING SUGGESTS THAT THE GUIDEWIRE SEPARATED DURING THE TIP SHAPING PROCESS, THE DISTAL PORTION OF THE SUBJECT DEVICE WAS NOT RETURNED, SO THIS COULD NOT BE DEFINITIVELY IDENTIFIED AS A CAUSE OR CONTRIBUTOR. AS A RESULT, A ROOT CAUSE OF UNDETERMINABLE WAS ASSIGNED.
ANALYSIS OF THE RETURNED DEVICE FOUND THAT THE DISTAL TIP OF THE GUIDEWIRE HAD DETACHED. THE DEVICE HAD NOT BEEN USED IN THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FATHOM GUIDEWIRE | WIRE, GUIDE, CATHETER | DQX | BOSTON SCIENTIFIC - SALT LAKE CITY | M001509100 | B20100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |