FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2181795 · Received July 27, 2011

Report

Report Number
2531779-2011-05374
Event Type
Malfunction
Date Received
July 27, 2011
Date of Event
June 30, 2011
Report Date
July 1, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE UP AND CONTRAST BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE CONTRAST BUTTON HAS NO EFFECT ON INSULIN DELIVERY FUNCTION. THE KEYPAD WAS REMOVED AND CONTAMINATION WAS OBSERVED UNDER ALL OF THE BUTTON CONTACTS.

Additional Manufacturer Narrative · 1

(B)(6). THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

A FAMILY MEMBER REPORTED THAT THE DOWN ARROW KEYPAD BUTTON IS INTERMITTENTLY UNRESPONSIVE. SHE CONFIRMED THAT THE KEYPAD IS INTACT. SHE STATED THAT THE PATIENT WEARS THE PUMP IN A PUMP SKIN, AND DOES NOT CLEAN THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 14 YR