FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM

MDR report key: 2181771 · Received July 27, 2011

Report

Report Number
3005075853-2011-03041
Event Type
Malfunction
Date Received
July 27, 2011
Date of Event
June 10, 2011
Report Date
June 13, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS OF DEVICE (A) FOUND THAT IT WAS RECEIVED WITH THE SHAFT BENT AT HANDLE ASSEMBLY AREA MAKING THE INSTRUMENT NON-FUNCTIONAL. NO FUNCTIONAL TEST COULD BE PERFORMED DUE TO THE CONDITION OF THE RETURNED DEVICE. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS. THE ANALYSIS RESULTS OF DEVICE (B) FOUND THAT IT WAS RETURNED EMPTY AND WITH THE LOCK OUT MECHANISM BROKEN AS IT WAS FIRED THROUGH. THE INSTRUMENT IS DESIGNED TO LOCK OUT AFTER ALL THE CLIPS HAVE BEEN FIRED; THEREFORE A POTENTIAL CAUSE OF THE CUSTOMER REPORTED EXPERIENCE IS THE FIRING OF ALL OF THE CLIPS AND THE INSTRUMENT "WOULD NOT FIRE" (ACTIVATION OF THE LOCK OUT MECHANISM). THE INSTRUMENT HAS AN ORANGE INDICATOR THAT APPEARS ON THE TOP OF THE HANDLE AS A REFERENCE FOR THE USER AS TO THE QUANTITY OF CLIPS REMAINING. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS. BATCH # H9020C, MFG DATE 01/11/2011; MFG DATE 12/11/2015. (B)(4) LOCKOUT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, ALTHOUGH THE FIRST CLIP COULD BE FIRED PROPERLY, THE CLIP BECAME NOT TO BE FED INTO THE JAW AND COULD NOT BE FIRED FROM SECOND FIRING. THE SAME EVENT OCCURRED IN THE SECOND DEVICE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1