FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2181768 · Received July 27, 2011

Report

Report Number
2531779-2011-05373
Event Type
Malfunction
Date Received
July 27, 2011
Report Date
July 1, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/14/2011 WITH THE FOLLOWING FINDINGS: INVESTIGATION CONFIRMED THAT THE AUDIO BOLUS IS INTERMITTENTLY RESPONSIVE AND DIFFICULT TO PRESS. EVALUATION REVEALED THAT THE BOLUS BUTTON PLUG IS MISALIGNED. ALL OTHER KEYPAD BUTTONS WERE FOUND TO BE RESPONDING APPROPRIATELY TO BUTTON PRESSES, AND THERE WERE NO OTHER KEYPAD DEFECTS FOUND ON INVESTIGATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE AUDIO BOLUS BUTTON HAS BEEN INTERMITTENTLY UNRESPONSIVE FOR THE PAST COUPLE OF WEEKS. HE STATED THAT HE WEARS THE PUMP IN A CASE ON HIS BELT, AND DENIED CLEANING THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 62 YR