ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-05372
- Event Type
- Malfunction
- Date Received
- July 27, 2011
- Report Date
- July 1, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE ISSUE WITH THE KEYPAD BUTTON MALFUNCTION WAS CONFIRMED THROUGH PRODUCT ANALYSIS. THE KEYPAD BUTTONS ARE INTERMITTENTLY RESPONDING TO USER INPUTS DURING TESTING. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER THE KEY CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THIS COMPLAINT IS BEING REPORTED AS A MALFUNCTION DUE TO THE ALLEGED KEYPAD ISSUE. REPORTEDLY, ALL KEYPAD BUTTONS ARE INTERMITTENTLY NOT RESPONDING. THE ISSUE WAS NOT RESOLVED WITH TRAINING. THERE WAS NO EVIDENCE OF PRODUCT MISUSE. THE ANIMAS PRODUCT IS BEING RETURNED FOR INVESTIGATION. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR |