HOMECHOICE
Report
- Report Number
- 1423500-2011-09848
- Event Type
- Malfunction
- Date Received
- July 27, 2011
- Date of Event
- June 25, 2011
- Report Date
- July 15, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). THE PAL EVALUATED THE DEVICE AND NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) FOUND IN THE DEVICE LOGS. THE PROBABLE CAUSE FOR THE IIPV WAS DETERMINED TO BE INSUFFICIENT DRAIN- FALSE EMPTY DETECT AT INITIAL DRAIN (I-DRAIN). DEVICE MET SPECIFICATIONS RELATIVE TO IIPV IN THE LOGS. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA.
(B)(4) THE DEVICE HAS BEEN RECEIVED AT BAXTER FOR EVALUATION. HOWEVER, THE EVALUATION IS NOT COMPLETE AT THIS TIME. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
DURING INITIAL ASSESSMENT OF A RETURNED HOMECHOICE MACHINE, A BAXTER TECHNICIAN FOUND AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) THAT WAS IDENTIFIED IN THE LOGS THAT OCCURRED ON DATE (B)(6) 2011 AT 23:46:36 WITH DRAIN VOLUME OF 3287 ML DURING CYCLE 1. THE FILL VOLUME IS 2000 ML. THIS EVENT MEETS OVERFILL CRITERIA. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR |