HOMECHOICEPRO
Report
- Report Number
- 1423500-2011-09854
- Event Type
- Malfunction
- Date Received
- July 27, 2011
- Date of Event
- June 28, 2011
- Report Date
- July 13, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). THE PAL EVALUATED THE DEVICE AND NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) FOUND IN THE DEVICE LOGS. THE CAUSE OF THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) WAS DETERMINED TO BE INSUFFICIENT DRAIN, ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. DEVICE MET SPECIFICATIONS RELATIVE TO IIPV IN THE LOGS. THERE WERE NO PROBLEMS FOUND DURING AN INTERNAL INSPECTION OF THE DEVICE THAT WOULD CONTRIBUTE TO THE IIPV. A REVIEW OF THE DEVICE SERVICE HISTORY FOR THIS PARTICULAR HOMECHOICE REVEALED THE PREVIOUS SWAP WAS UNRELATED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THIS ISSUE HAS BEEN ESCALATED TO CAPA.
(B)(4). THE DEVICE HAS BEEN RECEIVED AT BAXTER FOR EVALUATION. HOWEVER, THE EVALUATION IS NOT COMPLETE AT THIS TIME. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
DURING EVALUATION OF THE HOMECHOICE (HC) MACHINE, AN ADDITIONAL ISSUE OF AN INCREASED INTRAPERITONEAL VOLUME (IIPV) EVENT WAS FOUND IN THE PATIENT EVENT LOGS: OCCURRENCE DATE (B)(6) 2011 WITH DRAIN VOLUME OF 3587 MILLILITERS (ML) DURING CYCLE 4. THE LARGEST PRESCRIBED FILL VOLUME (LPFV) WAS 2200 ML. THIS MEETS THE CRITERIA FOR IIPV. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.