FDA Adverse Event Other Summary report: N

HOYER ELECTRIC LIFT

MDR report key: 2181530 · Received July 18, 2011

Report

Report Number
2182305-2011-00017
Event Type
Other
Date Received
July 18, 2011
Date of Event
May 16, 2011
Report Date
June 13, 2011
Manufacturer
APEX HEALTHCARE MFG INC
Product Code
FNG
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

IT WAS REPORTED TO MANUFACTURER, BY DSI, ONE OF THEIR CUSTOMERS (B)(6), PER FACILITIES CUSTOMER SERVICE CALL, (B)(6) REFERENCE AN INCIDENT THAT ALLEGEDLY INVOLVED THE LIFT. HE REPORTED THE FOLLOWING OCCURRED IN THE EVENING OF (B)(6) 2011; TWO C.N.A.'S WERE MOVING A RESIDENT FROM THE BED TO A WHEELCHAIR, WHEN PART OF THE LIFT CAME APART, AND THEY DROPPED THE RESIDENT INTO THE WHEELCHAIR. PRODUCT HAS BEEN TAKEN OUT OF SERVICE. PICTURES OF THE LIFT WERE SENT TO THE MANUFACTURER FOR REVIEW. PICTURES REVEALED ON THE 4 POINT CRADLE, THERE ARE 2 SWIVEL BOLTS / PINS IN THE HINGE, ONE ON EITHER SIDE OF THE CRADLE, WHERE THE STRAP HOOKS TO. ONE OF THE BOLTS SNAPPED OFF. AFTER CONVERSATION WITH THE (B)(6) NO INJURY TO THE RESIDENT WAS REPORTED DUE TO THE FAST RESPONSE OF THEIR AIDES; HOWEVER, THEY BELIEVE THIS COULD HAVE BECOME A SERIOUS ISSUE IF THE AIDES WERE NOT SO ATTENTIVE. (B)(4) AND (B)(4) ISSUED TO GET LIFT PARTS BACK FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOYER ELECTRIC LIFT PATIET LIFTER FNG APEX HEALTHCARE MFG INC PRESENCE W/4PTWSCLP

Patients

Seq Age Sex Outcome Treatment
1 UNK Other