S5 STOCKERT CENTRIFUGAL PUMP SYSTEM
Report
- Report Number
- 1718850-2011-00083
- Event Type
- Other
- Date Received
- July 15, 2011
- Date of Event
- April 7, 2011
- Report Date
- June 15, 2011
- Manufacturer
- SORIN GROUP DEUTSCHLAND
- Product Code
- DWA
- PMA / PMN Number
- K032213
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
SORIN GROUP (B)(4) MANUFACTURES THE CENTRIFUGAL PUMP. THIS MEDWATCH REPORT IS BEING FILED ON BEHALF OF SORIN GROUP (B)(4). SORIN GROUP (B)(4) RECEIVED A REPORT OF A SQUEALING NOISE FROM THE REVOLUTION PUMP HEAD AND PUMP DRIVER, WHICH RESULTED IN LOW FLOW. THE PUMP DRIVER WAS MANUALLY CRANKED TO WEAN THE PT FROM BYPASS, THEN THE PUMP HEAD WAS RESEATED INTO THE PUMP DRIVER AND THE CASE CONTINUED. THERE WAS NO REPORT OF PT INJURY. HOWEVER, FURTHER COMMUNICATION WITH THE PERFUSIONIST REVEALED THAT SUB-OPTIMAL HAND CRANKING WAS NOT ABLE TO GENERATE SUFFICIENT FLOW FOR APPROXIMATELY 4 MIN BEFORE ASSISTANCE WAS PROVIDED. THE SCP CENTRIFUGAL PUMP WAS RETURNED TO SORIN GROUP (B)(4) FOR EVALUATION. UPON RECEIPT, THE UNIT WAS SUBJECTED TO SIMULATED USE TESTING FOR APPROXIMATELY 12 HOURS AT HIGH AND LOW SPEED, WITH AND WITHOUT BACKPRESSURE. DURING INITIAL SIMULATED USE TESTING, THE PROBLEM WAS NOT DUPLICATED; THE DEVICE SUCCESSFULLY COMPLETED FUNCTIONAL TESTING. THE SCP PUMP PERFORMED AS EXPECTED. THE FLOW RATE WAS MAINTAINED AND NO ABNORMAL NOISE WAS PRODUCED AT ANY POINT THROUGH THE DURATION OF TESTING. THE REPORT OF HEARING A SQUEALING NOISE AND NOTICING THE FLOWS DROPPING OFF COULD NOT BE REPRODUCED AND THE CAUSE FOR THE CLINICIANS CONCERN COULD NOT BE DETERMINED. NO STRUCTURAL OR PERFORMANCE DEFECTS WERE FOUND. THE SCP PUMP HAS BEEN SENT TO SORIN GROUP (B)(4) FOR FURTHER EVALUATION. THE INVESTIGATION IS ON GOING. A FOLLOW UP REPORT WILL BE FILED WHEN THE INVESTIGATION IS COMPLETE AND RESULTS ARE AVAILABLE.
SORIN GROUP (B)(4) RECEIVED A REPORT OF A SQUEALING NOISE FROM THE REVOLUTION PUMP HEAD AND PUMP DRIVER, WHICH RESULTED IN LOW FLOW. THE PUMP DRIVER WAS MANUALLY CRANKED TO WEAN THE PT FROM BYPASS, THEN THE PUMP HEAD WAS RESEATED INTO THE PUMP DRIVER AND THE CASE CONTINUED. THERE WAS NO REPORT OF PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | S5 STOCKERT CENTRIFUGAL PUMP SYSTEM | PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE | DWA | SORIN GROUP DEUTSCHLAND | 60-00-50 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |