FDA Adverse Event
Other
Summary report: N
OLYMPUS SONOSURG SCISSORS
MDR report key: 2181514
·
Received May 25, 2011
Report
- Report Number
- 8010047-2011-00105
- Event Type
- Other
- Date Received
- May 25, 2011
- Date of Event
- March 18, 2011
- Report Date
- April 26, 2011
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORPORATION
- Product Code
- LFL
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
THE REFERENCED DEVICE WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP (OMSC) FOR EVAL. THE EVAL FOUND A BROKEN PIN, BENT GRASPING PORTION, AND SCRATCH MARK ON THE GRASPING PORTION OF THE DEVICE. OMSC ATTRIBUTED THE REPORTED PHENOMENON TO PHYSICAL DAMAGE. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.
Description of Event or Problem · 1
THE USER FACILITY REPORTED THAT DURING A NEPHRECTOMY, A DEVICE FRAGMENT FELL INTO THE PT. THE DEVICE FRAGMENT WAS REMOVED USING A GRASPING FORCEPS (MODEL UNK). THE PROCEDURE WAS COMPLETED AND THERE WAS NO PT INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OLYMPUS SONOSURG SCISSORS | SONOSURG SCISSORS | LFL | OLYMPUS MEDICAL SYSTEMS CORPORATION | T3105 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Other |