FDA Adverse Event Other Summary report: N

OLYMPUS SONOSURG SCISSORS

MDR report key: 2181514 · Received May 25, 2011

Report

Report Number
8010047-2011-00105
Event Type
Other
Date Received
May 25, 2011
Date of Event
March 18, 2011
Report Date
April 26, 2011
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORPORATION
Product Code
LFL
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE REFERENCED DEVICE WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP (OMSC) FOR EVAL. THE EVAL FOUND A BROKEN PIN, BENT GRASPING PORTION, AND SCRATCH MARK ON THE GRASPING PORTION OF THE DEVICE. OMSC ATTRIBUTED THE REPORTED PHENOMENON TO PHYSICAL DAMAGE. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT DURING A NEPHRECTOMY, A DEVICE FRAGMENT FELL INTO THE PT. THE DEVICE FRAGMENT WAS REMOVED USING A GRASPING FORCEPS (MODEL UNK). THE PROCEDURE WAS COMPLETED AND THERE WAS NO PT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OLYMPUS SONOSURG SCISSORS SONOSURG SCISSORS LFL OLYMPUS MEDICAL SYSTEMS CORPORATION T3105 NA

Patients

Seq Age Sex Outcome Treatment
1 62 YR Other