FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2181498 · Received July 7, 2011

Report

Report Number
1720753-2011-08824
Event Type
Malfunction
Date Received
July 7, 2011
Date of Event
June 28, 2011
Report Date
July 7, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE MEMORY BATTERY ON THE X-RAY CONTROLLER WAS REPLACED, AND FILAMENT CALIBRATION WAS PERFORMED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A FILAMENT ERROR MESSAGE DURING A CASE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1