FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 2181498
·
Received July 7, 2011
Report
- Report Number
- 1720753-2011-08824
- Event Type
- Malfunction
- Date Received
- July 7, 2011
- Date of Event
- June 28, 2011
- Report Date
- July 7, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE MEMORY BATTERY ON THE X-RAY CONTROLLER WAS REPLACED, AND FILAMENT CALIBRATION WAS PERFORMED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A FILAMENT ERROR MESSAGE DURING A CASE. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |