FDA Adverse Event Malfunction Summary report: N

6800

MDR report key: 2181492 · Received July 7, 2011

Report

Report Number
1720753-2011-08835
Event Type
Malfunction
Date Received
July 7, 2011
Date of Event
June 29, 2011
Report Date
July 7, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE REPORTED ISSUE WAS DUPLICATED. THE PRINTED CIRCUIT BOARDS AND THE CABLES WERE RESEATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED, THE SYSTEM WOULD NOT BOOT UP. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 6800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 6800

Patients

Seq Age Sex Outcome Treatment
1