NC QUANTUM APEX PTCA DILATATION CATHETER

MDR report key: 2181455 · Received July 27, 2011

Report

Report Number
2134265-2011-03029
Event Type
Malfunction
Date Received
July 27, 2011
Date of Event
July 4, 2011
Report Date
July 4, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATED BY MANUFACTURER: THE COMPLAINT DEVICE WAS NOT RETURNED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE, A BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED AND MODERATELY TORTUOUS DISTAL LEFT CIRCUMFLEX ARTERY. THE 15MM X 2.50MM NC QUANTUM APEX BALLOON WAS USED FOR POST-DILATATION FOLLOWING PARTIAL EXPANSION OF AN UNKNOWN STENT. THE NC QUANTUM APEX WAS INFLATED SEVERAL TIMES AT 18ATMS. DUE TO THE STENT NOT EXPANDING ENOUGH, THE NC QUANTUM APEX WAS INFLATED TO 20ATMS WHEN THE BALLOON RUPTURE OCCURRED. THE PROCEDURE WAS COMPLETED WITH ANOTHER MANUFACTURERS BALLOON DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NC QUANTUM APEX PTCA DILATATION CATHETER CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493912415250 13728617

Patients

Seq Age Sex Outcome Treatment
1