FDA Adverse Event Injury Summary report: N

TALENT ABDOMINAL STENT GRAFT

MDR report key: 2181397 · Received July 21, 2011

Report

Report Number
2953200-2011-01310
Event Type
Injury
Date Received
July 21, 2011
Date of Event
June 16, 2011
Report Date
June 22, 2011
Manufacturer
MEDTRONIC CARDIOVASCULAR IRELAND
Product Code
MIH
PMA / PMN Number
P070027
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION, RESULTS: (MIGRATION, ENDOLEAK). RESULTS AND CONCLUSIONS: (ANGULATED NECK).

Description of Event or Problem · 1

A TALENT ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM 16 MONTHS AGO. VESSEL MORPHOLOGY AT TIME OF IMPLANT WAS REPORTED AS THE AORTIC NECK HAD A 60 DEGREE ANGLE. IT WAS REPORTED THAT THERE WAS NO ENDOLEAK ON THE ANGIOGRAM TAKEN AFTER IMPLANT. IT WAS REPORTED THAT THE PATIENT WAS SEEN FOR A ROUTINE FOLLOW-UP, ONE MONTH AGO, AND THE STENT GRAFT HAD MIGRATED 1 TO 1.5 CM DISTALLY RESULTING IN A PROXIMAL TYPE 1 ENDOLEAK. THE MIGRATION WAS ATTRIBUTED TO THE NECK ANGULATION. THE PATIENT WAS TREATED WITH TWO ENDURANT CUFFS THREE WEEKS LATER. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IF FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT ABDOMINAL STENT GRAFT MIH MEDTRONIC CARDIOVASCULAR IRELAND NA V00431067

Patients

Seq Age Sex Outcome Treatment
1 80 YR Required Intervention