TALENT ABDOMINAL STENT GRAFT
Report
- Report Number
- 2953200-2011-01310
- Event Type
- Injury
- Date Received
- July 21, 2011
- Date of Event
- June 16, 2011
- Report Date
- June 22, 2011
- Manufacturer
- MEDTRONIC CARDIOVASCULAR IRELAND
- Product Code
- MIH
- PMA / PMN Number
- P070027
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVALUATION, RESULTS: (MIGRATION, ENDOLEAK). RESULTS AND CONCLUSIONS: (ANGULATED NECK).
A TALENT ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM 16 MONTHS AGO. VESSEL MORPHOLOGY AT TIME OF IMPLANT WAS REPORTED AS THE AORTIC NECK HAD A 60 DEGREE ANGLE. IT WAS REPORTED THAT THERE WAS NO ENDOLEAK ON THE ANGIOGRAM TAKEN AFTER IMPLANT. IT WAS REPORTED THAT THE PATIENT WAS SEEN FOR A ROUTINE FOLLOW-UP, ONE MONTH AGO, AND THE STENT GRAFT HAD MIGRATED 1 TO 1.5 CM DISTALLY RESULTING IN A PROXIMAL TYPE 1 ENDOLEAK. THE MIGRATION WAS ATTRIBUTED TO THE NECK ANGULATION. THE PATIENT WAS TREATED WITH TWO ENDURANT CUFFS THREE WEEKS LATER. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IF FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TALENT ABDOMINAL STENT GRAFT | MIH | MEDTRONIC CARDIOVASCULAR IRELAND | NA | V00431067 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Required Intervention |