ATS 1200 TOURNIQUET SYSTEM
Report
- Report Number
- 1526350-2011-00142
- Event Type
- Malfunction
- Date Received
- July 8, 2011
- Date of Event
- June 1, 2011
- Report Date
- June 9, 2011
- Manufacturer
- ZIMMER SURGICAL
- Product Code
- KCY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NM, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS RETURNED TO THE MFR FOR REPAIR. THE SVC RECORDS INDICATE THAT THE DEVICE IS 8 YEARS OLD AND HAS BEEN RETURNED FOR REPAIR PREVIOUSLY. THE PUMP ON THE ATS 1200 IS SUCH THAT IT WILL MAINTAIN PRESSURE IN ALL BUT A CATASTROPHIC PRESSURE LOSS, SUCH AS A LARGE CUFF RUPTURE. THIS DOES NOT APPEAR TO BE THE CASE WITH THIS COMPLAINT. CLINICAL F/U DID NOT INDICATE AN AUDIBLE ALARM SOUNDED, WHICH WOULD BE THE CASE IF A GREATER THAN 15MMHG OF PRESSURE WAS LOST. THE HOSPITAL STAFF ALSO STATED "THE CUFF WAS MOST LIKELY PUT ON IMPROPERLY BY A TRAVELER STAFF." THIS IS A MORE LIKELY SCENARIO THAT THE CUFF WOULD MOVE, CAUSING A TEMPORARY PRESSURE LOSS, THEN INFLATING TO MAINTAIN THE SET PRESSURE. THE CUFF WAS NOT RETURNED FOR EVALUATION. THE DEVICE WAS SERVICED AND RETURNED TO THE CUSTOMER.
IT WAS REPORTED THAT THE ZIMMER ATS 1200 "LET GO OF PRESSURE DURING CASE." ADDITIONAL CLINICAL F/U WITH THE HOSPITAL INDICATED THAT "THE TOURNIQUET CUFF MOST LIKELY WAS PUT ON IMPROPERLY BY TRAVELER STAFF WHO WERE NO LONGER AT THE FACILITY." THERE WAS REPORT OF SOME BLOOD IN THE FIELD BUT NOT MUCH. THERE WAS NO REPORTED HARM TO THE PT OR ADDITIONAL INTERVENTION REQUIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATS 1200 TOURNIQUET SYSTEM | ATS 1200 TOURNIQUET SYSTEM | KCY | ZIMMER SURGICAL | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |