ENDURANT ABDOMINAL STENT GRAFT SYSTEM
Report
- Report Number
- 2953200-2011-01328
- Event Type
- Injury
- Date Received
- July 22, 2011
- Date of Event
- June 24, 2011
- Report Date
- June 24, 2011
- Manufacturer
- MEDTRONIC CARDIOVASCULAR IRELAND
- Product Code
- MIH
- PMA / PMN Number
- P100021
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVALUATION, RESULTS: (IMPLANTATION INTO A PATIENT WITH A SEVERELY ANGULATED NECK), (ENDOLEAK, MIGRATION). CONCLUSION: (ANGULATION OF NECK).
AN ENDURANT BIFURCATED STENT GRAFT WAS IMPLANTED IN A PATIENT FOR A SLOW GROWING, GREATER THAN 5 CM IN DIAMETER ABDOMINAL ANEURYSM, APPROXIMATELY 18 WEEKS AGO. VESSEL MORPHOLOGY WAS REPORTED AS THE AORTIC NECK WAS 24-26 MM IN DIAMETER FOR ABOUT 1 CM IN LENGTH, THEN HAD A SEVERE BEND OF 60-70 DEGREES AND ENLARGED TO 30 MM IN DIAMETER WITH MILD CALCIFICATION AND THROMBUS. THE ILIAC ARTERIES WERE NON-TORTUOUS BUT VERY HARD AND CALCIFIED. THE FEMORAL ARTERIES WERE 3 MM IN DIAMETER AND THERE WAS A PATENT IMA. COMORBIDITIES INCLUDES HIGH CREATININE AND OSTEOPOROSIS, COMPRESSED VERTEBRAE/SPONDYLOSIS, AND LEFT-RIGHT CURVATURE OF THE SPINE. THE AORTA FOLLOWED THE CURVATURE OF THE SPINE. THE PATIENT WAS NOT AN OPEN SURGICAL REPAIR CANDIDATE. THE ENDURANT BIFURCATED STENT GRAFT AND CONTRALATERAL LIMB WERE IMPLANTED WITHOUT ISSUES, WITH PROXIMAL PLACEMENT RIGHT AT THE RENALS. AN ULTRASOUND WAS PERFORMED THREE OR FOUR MONTHS AFTER IMPLANT, RATHER THAN A FOLLOW UP CT, DUE TO THE PATIENT'S INABILITY TO HANDLE CONTRAST MEDIA. THE ULTRASOUND SHOWED WHAT MAY HAVE BEEN AN ENDOLEAK AND SAC ENLARGEMENT. AN ANGIOGRAM CONFIRMED A PROXIMAL TYPE I ENDOLEAK, AS THE STENT GRAFT HAD MIGRATED APPROXIMATELY 1 CM DISTALLY, WITH THE SUPRARENAL STRUTS BELOW THE LOWEST RENAL. THE AAA WAS FOUND TO HAVE ENLARGED TO 6 CM. AN ENDURANT 32X32X49 CUFF WAS PLACED PROXIMALLY AT THE RENALS AND BALLOONED WITH ANOTHER MANUFACTURER'S BALLOON WHICH RESOLVED THE PROXIMAL TYPE I ENDOLEAK. SEVERAL MORE SCANS FROM DIFFERENT ANGLES CONFIRMED THE RESOLUTION OF THE ENDOLEAK. THE MIGRATION WAS LIKELY RELATED TO THE CHALLENGING NECK ANATOMY. NO CLINICAL SEQUELAE WERE REPORTED, AND THE PATIENT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDURANT ABDOMINAL STENT GRAFT SYSTEM | MIH | MEDTRONIC CARDIOVASCULAR IRELAND | NA | V00804844 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Required Intervention |