FDA Adverse Event Injury Summary report: N

ENDURANT ABDOMINAL STENT GRAFT SYSTEM

MDR report key: 2181217 · Received July 22, 2011

Report

Report Number
2953200-2011-01328
Event Type
Injury
Date Received
July 22, 2011
Date of Event
June 24, 2011
Report Date
June 24, 2011
Manufacturer
MEDTRONIC CARDIOVASCULAR IRELAND
Product Code
MIH
PMA / PMN Number
P100021
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION, RESULTS: (IMPLANTATION INTO A PATIENT WITH A SEVERELY ANGULATED NECK), (ENDOLEAK, MIGRATION). CONCLUSION: (ANGULATION OF NECK).

Description of Event or Problem · 1

AN ENDURANT BIFURCATED STENT GRAFT WAS IMPLANTED IN A PATIENT FOR A SLOW GROWING, GREATER THAN 5 CM IN DIAMETER ABDOMINAL ANEURYSM, APPROXIMATELY 18 WEEKS AGO. VESSEL MORPHOLOGY WAS REPORTED AS THE AORTIC NECK WAS 24-26 MM IN DIAMETER FOR ABOUT 1 CM IN LENGTH, THEN HAD A SEVERE BEND OF 60-70 DEGREES AND ENLARGED TO 30 MM IN DIAMETER WITH MILD CALCIFICATION AND THROMBUS. THE ILIAC ARTERIES WERE NON-TORTUOUS BUT VERY HARD AND CALCIFIED. THE FEMORAL ARTERIES WERE 3 MM IN DIAMETER AND THERE WAS A PATENT IMA. COMORBIDITIES INCLUDES HIGH CREATININE AND OSTEOPOROSIS, COMPRESSED VERTEBRAE/SPONDYLOSIS, AND LEFT-RIGHT CURVATURE OF THE SPINE. THE AORTA FOLLOWED THE CURVATURE OF THE SPINE. THE PATIENT WAS NOT AN OPEN SURGICAL REPAIR CANDIDATE. THE ENDURANT BIFURCATED STENT GRAFT AND CONTRALATERAL LIMB WERE IMPLANTED WITHOUT ISSUES, WITH PROXIMAL PLACEMENT RIGHT AT THE RENALS. AN ULTRASOUND WAS PERFORMED THREE OR FOUR MONTHS AFTER IMPLANT, RATHER THAN A FOLLOW UP CT, DUE TO THE PATIENT'S INABILITY TO HANDLE CONTRAST MEDIA. THE ULTRASOUND SHOWED WHAT MAY HAVE BEEN AN ENDOLEAK AND SAC ENLARGEMENT. AN ANGIOGRAM CONFIRMED A PROXIMAL TYPE I ENDOLEAK, AS THE STENT GRAFT HAD MIGRATED APPROXIMATELY 1 CM DISTALLY, WITH THE SUPRARENAL STRUTS BELOW THE LOWEST RENAL. THE AAA WAS FOUND TO HAVE ENLARGED TO 6 CM. AN ENDURANT 32X32X49 CUFF WAS PLACED PROXIMALLY AT THE RENALS AND BALLOONED WITH ANOTHER MANUFACTURER'S BALLOON WHICH RESOLVED THE PROXIMAL TYPE I ENDOLEAK. SEVERAL MORE SCANS FROM DIFFERENT ANGLES CONFIRMED THE RESOLUTION OF THE ENDOLEAK. THE MIGRATION WAS LIKELY RELATED TO THE CHALLENGING NECK ANATOMY. NO CLINICAL SEQUELAE WERE REPORTED, AND THE PATIENT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDURANT ABDOMINAL STENT GRAFT SYSTEM MIH MEDTRONIC CARDIOVASCULAR IRELAND NA V00804844

Patients

Seq Age Sex Outcome Treatment
1 71 YR Required Intervention