ENDO GIA ULTRA
Report
- Report Number
- 2647580-2025-01122
- Event Type
- Malfunction
- Date Received
- April 10, 2025
- Date of Event
- March 15, 2025
- Report Date
- June 19, 2025
- Manufacturer
- US SURGICAL PUERTO RICO
- Product Code
- GDW
- PMA / PMN Number
- K111825
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
D10 CONCOMITANT PRODUCTS: BOX00662V1, BOX APPENDECTOMY KIT 3.5MM BLUE, (LOT # 230710570); BOX00662V1, BOX APPENDECTOMY KIT 3.5MM BLUE, (LOT # 230742640); 30419, 030419 ENDO GIAII 30 3.5MM DLU X6, (LOT # UNKNOWN); EGIAUSTND, EGIAUSTND ENDOGIA ULTRA UNIV STD STAP, (LOT # UNKNOWN); 30419, 030419 ENDO GIAII 30 3.5MM DLU X6, (LOT # UNKNOWN). MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
ADDITIONAL INFORMATION: D9, E1 (STREET 1, CITY, POSTAL CODE AND PHONE NUMBER), E4 (EMAIL), G3, H3, H6. H3 EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. THE EVALUATION FOUND NO POTENTIALLY CONTRIBUTING FACTORS, AND THE SAMPLE MET ALL RELATED SPECIFICATIONS. IT WAS REPORTED THAT THE JAWS DID NOT OPEN. THE REPORTED ISSUE COULD NOT BE CONFIRMED. THE MOST LIKELY CAUSE COULD NOT BE IDENTIFIED BECAUSE NO RELATED PROBLEM WAS DETECTED WITH THE DEVICE. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
ACCORDING TO THE REPORTER, IN LAPAROSCOPIC APPENDECTOMY, DURING SET UP PRIOR TO FIRING, THE JAWS OF THE STAPLER COULD NOT BE OPENED. A SECOND STAPLER WAS USED AND THE SECOND STAPLER HAD FIRING ISSUES. THE SECOND STAPLER FIRED BUT GOT JAMMED. A THIRD STAPLER WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1422014 | ENDO GIA ULTRA | STAPLE, IMPLANTABLE | GDW | US SURGICAL PUERTO RICO | EGIAUSTND |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | SEE H11 |