FDA Adverse Event Malfunction Summary report: N

ROTATING MULTIPLE CLIP APPLIER

MDR report key: 2181076 · Received July 27, 2011

Report

Report Number
3005075853-2011-03035
Event Type
Malfunction
Date Received
July 27, 2011
Date of Event
July 7, 2011
Report Date
July 11, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS OF THE DEVICE FOUND THAT IT WAS RECEIVED WITH THE TOP AND LOWER SHROUD BROKEN CAUSING THAT THE CLIPS TO BE EJECTED. HOWEVER, IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THIS CONDITION. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TO A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, A SCRUB NURSE FIRED THE DEVICE TO CHECK ITS FUNCTION BEFORE USE ON THE PATIENT, AND A CLIP EJECTED. THE EVENT OCCURRED SEVERAL TIMES, AND THEN A DEFORMED CLIP JAMMED IN THE JAWS. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ROTATING MULTIPLE CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1