FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2181051 · Received July 27, 2011

Report

Report Number
2531779-2011-05355
Event Type
Malfunction
Date Received
July 27, 2011
Date of Event
June 30, 2011
Report Date
June 30, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE OK BUTTON WAS FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES; NONE OF THE KEYPAD BUTTONS HAD NORMAL SPRING BACK. THE KEYPAD WAS REMOVED AND CORROSION WAS OBSERVED UNDER THE BUTTON CONTACTS.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE OK KEYPAD BUTTON BECAME UNRESPONSIVE WHILE ATTEMPTING TO PROGRAM A BOLUS. SHE STATED THAT THE BUTTON FEELS LIKE "A BUBBLE". SHE CONFIRMED THAT THE KEYPAD IS INTACT. THE PATIENT STATED THAT SHE WEARS THE PUMP IN HER BRA, AND DOES NOT CLEAN THE PUMP.