Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REPRESENTATIVE REPLACED THE CMOS BATTERY, AND THE X-RAY SWITCH, AND CHECKED THE VOLTAGES. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD LOCK UP AND CONTINUE TO PRODUCE FLUOROSCOPIC X-RAY BEYOND DEMAND. NO PATIENT INJURY WAS REPORTED.