FDA Adverse Event Malfunction Summary report: N

6600

MDR report key: 2181014 · Received July 7, 2011

Report

Report Number
1720753-2011-08805
Event Type
Malfunction
Date Received
July 7, 2011
Date of Event
June 22, 2011
Report Date
July 7, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REPRESENTATIVE REPLACED THE CMOS BATTERY, AND THE X-RAY SWITCH, AND CHECKED THE VOLTAGES. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD LOCK UP AND CONTINUE TO PRODUCE FLUOROSCOPIC X-RAY BEYOND DEMAND. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 6600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 6600

Patients

Seq Age Sex Outcome Treatment
1