FDA Adverse Event Malfunction Summary report: N

UNICEL DXC 600 SYNCHRON SYSTEM

MDR report key: 2180991 · Received July 1, 2011

Report

Report Number
2050012-2011-02757
Event Type
Malfunction
Date Received
July 1, 2011
Date of Event
February 25, 2010
Report Date
March 3, 2010
Manufacturer
BECKMAN COULTER, INC.
Product Code
CGA
PMA / PMN Number
K042291
Removal / Correction Number
Z-2388-2010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) VISITED THE FACILITY AND EXAMINED THE SYSTEM. THE FSE NOTED THAT THE SAMPLE WHEEL COVER WAS NOT PROPERLY PLACED ON THE SYSTEM. THE SAMPLE WHEEL COVER WAS PROPERLY SEATED. THE CUSTOMER WAS PREVIOUSLY ALERTED TO THIS SCENARIO. THE ROOT CAUSE OF THIS EVENT APPEARS TO BE SAMPLE WHEEL COVER MISALIGNMENT CONTRIBUTING TO IMPROPER MODULAR CHEMISTRY PROBE HEIGHT CLEARANCE. THIS IS ONE OF NINE MEDWATCH REPORTS BEING SUBMITTED AS THE CUSTOMER PROVIDED DATA FOR NINE INDIVIDUAL PT EVENTS. REFERENCE THE MDR NUMBERS BELOW FOR ALL ASSOCIATED EVENTS: MDR# 2050012-2011-02687, 02750, 02751, 02752, 02753, 02754, 02755, 02756. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW OF COMPLAINTS CONDUCTED BETWEEN JANUARY 1, 2008 AND OCTOBER 23, 2010 FOR ADD'L REPORTABLE EVENTS.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT ERRONEOUSLY LOW GLUCOSE RESULTS WERE GENERATED BY THE UNICEL DXC 600 SYNCHRON SYSTEM. THE SYSTEM GENERATED CRITICAL RERUN FEATURE WAS TRIGGERED AND ALSO GENERATED A LOW RESULT. THE INITIAL RESULT WAS REPORTED OUT OF THE LABORATORY. THE SAMPLE WAS RETESTED ON THE SAME SYSTEM AND ANOTHER DXC SYSTEM. THE SAME SYSTEM YIELDED A SIMILAR LOW RESULT AND THE OTHER SYSTEM YIELDED A RESULT WITHIN EXPECTATIONS. AMENDED RESULTS WERE REPORTED. QUALITY CONTROLS WERE WITHIN SPECIFICATION PRIOR TO AND AFTER THE EVENT. IT IS NOT KNOWN IF THERE WAS ANY CHANGE TO THE PT'S CARE OR TREATMENT. THERE IS NO REPORT OF ANY SERIOUS INJURY OR NEED FOR MEDICAL INTERVENTION TO PREVENT OR PRECLUDE SERIOUS INJURY RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL DXC 600 SYNCHRON SYSTEM CGA BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 48 YR