ENDURANT ILIAC STENT GRAFT
Report
- Report Number
- 2953200-2011-01314
- Event Type
- Injury
- Date Received
- July 22, 2011
- Date of Event
- June 23, 2011
- Report Date
- June 23, 2011
- Manufacturer
- MEDTRONIC CARDIOVASCULAR IRELAND
- Product Code
- MIH
- PMA / PMN Number
- P100021
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVAL RESULTS AND CONCLUSION: ENDOLEAK. AGGRESSIVE BALLOONING, TREATMENT OF ONLY AN ILIAC ANEURYSM.
AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 34MM DIAMETER RIGHT COMMON ILIAC ANEURYSM. VESSEL MORPHOLOGY WAS REPORTED AS UNREMARKABLE EXCEPT FOR A SMALL AREA OF FOCAL STENOSIS ON THE RIGHT (CONTRALATERAL) SIDE. THE BIFURCATED STENT GRAFT WAS IMPLANTED ON THE LEFT SIDE, FOLLOWED BY THE MAIN CONTRALATERAL LIMB (REF MFR #2953200-2011-01314) AND A (REF MFR # 2953200-2011-01315) ON THE RIGHT SIDE, WITH A 3-4CM OVERLAP BETWEEN THESE TWO DEVICES. ON THE FINAL ANGIOGRAM RUN THERE WAS A SMALL TYPE 3 ENDOLEAK COMING FROM THE JUNCTION BETWEEN THE TWO DEVICES. THIS AREA WAS BALLOONED VERY AGGRESSIVELY WITH A RELIANT BALLOON USING A 35 CC SYRINGE. IT IS UNK HOW MUCH FLUID WAS INJECTED, BUT THE INJECTION WAS RAPID AND THE BALLOON ELONGATED DURING THE PROCESS AND RUPTURED (REF MFR #564249 2953200-2011-01316). THERE WAS NO FABRIC TEAR OR VESSEL RUPTURE; HOWEVER, THE ENDOLEAK DID NOT RESOLVE. AN ILIAC (B)(4) WAS USED TO RELINE THE JUNCTION AREA AND SUCCESSFULLY RESOLVED THE ENDOLEAK. NO CLINICAL SEQUELAE WAS REPORTED AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDURANT ILIAC STENT GRAFT | MIH | MEDTRONIC CARDIOVASCULAR IRELAND | NA | V00855609 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Required Intervention |