FDA Adverse Event
Malfunction
Summary report: N
CONSTELLATION WITH LASER
MDR report key: 2180945
·
Received July 7, 2011
Report
- Report Number
- 2028159-2011-00760
- Event Type
- Malfunction
- Date Received
- July 7, 2011
- Date of Event
- June 2, 2011
- Report Date
- June 7, 2011
- Manufacturer
- ALCON - IRVINE TECHNOLOGY CENTER
- Product Code
- HQC
- PMA / PMN Number
- K063583
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE COMPANY REP EXAMINED THE SYSTEM AND REPLACED THE HOST MODULE AND THE POWER CONTROLLER PRINTED CIRCUIT BOARD (PCB). THE SYSTEM WAS THEN TESTED AND MET SPECS. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFO BECOMES AVAILABLE. (B)(4).
Description of Event or Problem · 1
A CUSTOMER REPORTED THAT THE UNIT TURNED OFF UNEXPECTEDLY DURING A PROCEDURE. THE CASE WAS COMPLETED AFTER THE FACILITY PLUGGED THE UNIT INTO A DIFFERENT POWER OUTLET. THERE WAS NO PT HARM REPORTED. ADD'L INFO HAS BEEN REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONSTELLATION WITH LASER | UNIT, PHACOFRAGMENTATION | HQC | ALCON - IRVINE TECHNOLOGY CENTER | CONSTELLATION | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |