FDA Adverse Event Malfunction Summary report: N

CONSTELLATION WITH LASER

MDR report key: 2180945 · Received July 7, 2011

Report

Report Number
2028159-2011-00760
Event Type
Malfunction
Date Received
July 7, 2011
Date of Event
June 2, 2011
Report Date
June 7, 2011
Manufacturer
ALCON - IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K063583
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REP EXAMINED THE SYSTEM AND REPLACED THE HOST MODULE AND THE POWER CONTROLLER PRINTED CIRCUIT BOARD (PCB). THE SYSTEM WAS THEN TESTED AND MET SPECS. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFO BECOMES AVAILABLE. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT THE UNIT TURNED OFF UNEXPECTEDLY DURING A PROCEDURE. THE CASE WAS COMPLETED AFTER THE FACILITY PLUGGED THE UNIT INTO A DIFFERENT POWER OUTLET. THERE WAS NO PT HARM REPORTED. ADD'L INFO HAS BEEN REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONSTELLATION WITH LASER UNIT, PHACOFRAGMENTATION HQC ALCON - IRVINE TECHNOLOGY CENTER CONSTELLATION NA

Patients

Seq Age Sex Outcome Treatment
1