CESAR-OMCP-VISUB (HM3000)
MDR report key: 2180727
·
Received May 5, 2011
Report
- Report Number
- 3003768277-2011-00363
- Event Type
- Malfunction
- Date Received
- May 5, 2011
- Report Date
- April 14, 2011
- Manufacturer
- PHILIPS HEALTHCARE
- Product Code
- IZF
- PMA / PMN Number
- K925302
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE FIELD SVC ENGINEER (FSE) WAS DISPATCHED TO THE SITE AND FOUND A PROBLEM IN THE VIEWING SUBSYSTEM CAUSED BY THE TEMPORAL PROCESSING BOARD (TPB) BOARD. THE FSE RECOMMENDED REPLACING THIS TPB BOARD, THE CUSTOMER HAD BROUGHT THE BOARD AND HAD THE FSE REPLACE IT. THE SYSTEM WORKS CORRECTLY AND IS BACK IN OPERATIONS. NO FURTHER ACTIONS REQUIRED. NO RISK FOR THE PT, USER OR BYSTANDERS.
Description of Event or Problem · 1
THE EQUIPMENT FAILS INTERMITTENTLY.