VASOVIEW HEMOPRO 2 EVH SYSTEM
Report
- Report Number
- 2242352-2011-00822
- Event Type
- Malfunction
- Date Received
- July 7, 2011
- Date of Event
- June 15, 2011
- Report Date
- June 15, 2011
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- GEI
- PMA / PMN Number
- K101274
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PHYSICIAN
Narratives
INVESTIGATION: VISUAL INSPECTION REVEALED THAT THE JAWS HAD NO NON CONFORMITIES AND SOME SIGNS OF CLINICAL USAGE. THERE WAS SOME EVIDENCE OF BLOOD. RESISTANCE AND JAW FORCE MEASUREMENTS PASSED SPECS. A FUNCTIONAL TEST EVALUATING THE THERMAL SHUT-DOWN PERFORMANCE OF THE UNIT WAS PERFORMED ON THE RETURNED DEVICE USING A REFERENCE POWER SUPPLY AND CABLE. THE DEVICE PERFORMED ACCORDING TO SPECS DURING THE DEVICE ACTIVATION. BASED UPON THIS OBSERVATION, THE REPORTED COMPLAINT FOR "INTERMITTENT POWER" COULD NOT BE CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VESSEL HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO 2 DID NOT WORK, "THE USER WAITED 4 X 30 SEC FOR IT TO RESET BUT IT STILL NEVER WORKED CORRECTLY, ALSO, IT SPARKED WHEN IT WAS TESTED WITH A WET LAP, BUT THE USER DECIDED TO USE IT ANYWAY." THE REPORTER CLARIFIED THAT THE DEVICE HAD ONLY BEEN USED MINIMALLY, AND IT KEPT SHUTTING DOWN. A NEW KIT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PT EFFECTS. THE PRODUCT WAS RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VASOVIEW HEMOPRO 2 EVH SYSTEM | ENDOSCOPIC VESSEL HARVESTING SYSTEM | GEI | MAQUET CARDIOVASCULAR, LLC | VH-4000 | 25035991 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |