LEGEND FOOTED ATTACHMENT
Report
- Report Number
- 1625507-2011-00099
- Event Type
- Malfunction
- Date Received
- July 27, 2011
- Report Date
- June 28, 2011
- Manufacturer
- MDT POWERED SURGICAL SOLUTIONS
- Product Code
- HBB
- PMA / PMN Number
- K020069
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
REPORT CONFIRMED. EVALUATION DETERMINED THAT THE FOOTED PORTION WAS DAMAGED BY TOOL CONTACT. ON FOLLOW-UP, IT WAS CONFIRMED THAT THERE WAS NO PATIENT IMPACT. THE USER MANUAL CONTAINS THE FOLLOWING WARNING "THE ATTACHMENT SHOULD NOT BE USED IF ANY PART OF THE ATTACHMENT APPEARS TO BE BENT, LOOSE, MISSING, OR DAMAGED. EXCESSIVE PRESSURE OR IMPROPER HANDLING, SUCH AS BENDING OR PRYING, OF THE ATTACHMENT OR DISSECTING TOOL MAY CAUSE INJURY TO THE PATIENT, OPERATOR AND/OR OPERATING ROOM STAFF." OUR RECOMMENDATION FOR FACTORY SERVICE IS BASED ON THE FACILITY'S USAGE; HOWEVER IT SHOULD NOT EXCEED 24 MONTHS. THIS DEVICE HAS BEEN IN USE FOR 32 MONTHS WITHOUT ANY RECORD OF FACTORY SERVICE.
DEVICE RETURNED FOR NON-SPECIFIED REPAIR. NO PATIENT IMPACT REPORTED. REPAIR ESCALATED TO COMPLAINT ON DECONTAMINATION DUE TO THE ATTACHMENT BEING DAMAGED BY TOOL CONTACT. ON FOLLOW-UP, IT WAS CONFIRMED THAT THERE WAS NO PATIENT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LEGEND FOOTED ATTACHMENT | MOTOR, DRILL, PNEUMATIC | HBB | MDT POWERED SURGICAL SOLUTIONS | N/A | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |