FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2180588 · Received July 27, 2011

Report

Report Number
2531779-2011-05348
Event Type
Malfunction
Date Received
July 27, 2011
Report Date
June 29, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/14/2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION CONFIRMED THAT THE UP ARROW AND DOWN ARROW KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE. THE BUTTONS REQUIRE MULTIPLE PRESSES WITH EXCESSIVE FORCE TO OBTAIN A RESPONSE. THERE WAS EVIDENCE OF CONTAMINATION FOUND UNDER ALL KEY CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE UP ARROW AND DOWN ARROW KEYPAD BUTTONS HAVE BEEN INTERMITTENTLY UNRESPONSIVE FOR ABOUT A MONTH. SHE CONFIRMED THAT THE KEYPAD IS INTACT. SHE STATED THAT SHE WEARS THE PUMP IN HER POCKET, AND OCCASIONALLY CLEANS THE PUMP AROUND THE BATTERY COMPARTMENT THREADS WITH A DRY Q-TIP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 51 YR