FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 2180548 · Received July 27, 2011

Report

Report Number
2050012-2011-03750
Event Type
Malfunction
Date Received
July 27, 2011
Date of Event
June 26, 2011
Report Date
June 26, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

BASED ON CUSTOMER FEEDBACK, BEC HAS OPENED CAPA (B)(4) TO INVESTIGATE THE ROOT CAUSE OF A REAGENT SYRINGE COMING LOOSE FROM ITS CONNECTION TO A T-VALVE.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BECKMAN COULTER, INC. (BEC) TO REPORT ERRONEOUSLY LOW RESULTS FOR IRON, GLUCOSE, AND TRANSFERRING BETWEEN TWO (2) PATIENTS ON THEIR UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER. THE ERRONEOUSLY LOW PATIENT SAMPLE RESULTS WERE REPORTED OUT OF THE LABORATORY. THE CUSTOMER REPORTED THAT THERE WAS NO IMPACT TO PATIENT TREATMENT. THE CUSTOMER PROVIDED DATA PERTAINING TO THE ERRONEOUS RESULTS REPORTED OUT OF THE LABORATORY FOR IRON AND GLUCOSE. NO SPECIFIC DATA WAS PROVIDED BY THE CUSTOMER FOR THE ERRONEOUS TRANSFERRIN RESULTS. THE PATIENT SAMPLES WERE RE-ASSAYED ON A SECOND ANALYZER AND AMENDED REPORTS WERE GENERATED TO CAPTURE THESE REPEAT RESULTS. THE CUSTOMER OBSERVED LIQUID ON THE COVER BELOW THE PROBE WASH COLLAR. THE CUSTOMER DID NOT DETERMINE THE CAUSE OF THE LEAK AND HAD REQUESTED SERVICE. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER'S SITE ON (B)(4) 2011. THE FSE OBSERVED THAT THE 3-WAY VALVE FOR THE SAMPLE PROBE WAS LEAKING. THE FSE REPLACED THE 3-WAY VALVE AND THEN RAN QUALITY CONTROLS WITH NO ERRORS OBSERVED. THE ANALYZER REPAIR WAS VERIFIED PER ESTABLISHED PROCEDURES. THE ROOT CAUSE OF THE ERRONEOUS RESULTS FOR IRON, GLUCOSE, AND TRANSFERRING IS ATTRIBUTED TO THE LEAKING 3-WAY VALVE FOR THE SAMPLE PROBE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER JJE BECKMAN COULTER, INC. NA

Patients

Seq Age Sex Outcome Treatment
1