FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2180519 · Received July 27, 2011

Report

Report Number
2531779-2011-05293
Event Type
Malfunction
Date Received
July 27, 2011
Report Date
June 28, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE KEYPAD BUTTONS BEGAN STICKING AND OVERLY-RESPONSIVE ONE TO TWO MONTHS AGO. SHE STATED THAT THE BUTTONS INTERMITTENTLY RESPOND TO PRESSES, AND CONTINUE TO SCROLL AFTER PRESSING. SHE NOTED THAT THE BUTTONS FEEL FLAT, AND CONFIRMED THAT THERE WERE NO HOLES OR TEARS IN THE KEYPAD. THE PATIENT STATED THAT SHE WEARS THE PUMP UNDER HER CLOTHES AGAINST THE BODY, AND DOES NOT CLEAN THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 26 YR