FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2180518 · Received July 27, 2011

Report

Report Number
2531779-2011-05340
Event Type
Malfunction
Date Received
July 27, 2011
Date of Event
June 26, 2011
Report Date
June 26, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS. REVIEW OF THE PUMP HISTORY VERIFIED LOSS OF PRIME WARNINGS DURING THE LOAD STEP. THE PUMP WAS REWOUND, LOADED, AND PRIMED WITH NO ALARMS DURING INVESTIGATION. THE PUMP WAS EXERCISED WITH NO ALARMS OCCURRING. THE FORCE SENSOR CALIBRATION READING WAS OUT OF SPECIFICATION HOWEVER THE RESISTANCE READING WAS WITHIN SPECIFICATIONS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PATIENT CALLED TO REPORT THAT HE CHANGED HIS INFUSION SET AND WHEN HE PRIMED HIS PUMP HE ONLY PRIMED 9 UNITS. HE PREVIOUSLY PRIMED 21.6 UNITS ON (B)(6) AND 21.8 UNITS ON (B)(6) BUT ON (B)(6) HE ONLY PRIMED 9 UNITS TO SEE 5 DROPS OF INSULIN COMING OUT OF THE END OF THE TUBING. THE 9 UNITS IS AN UNEXPECTEDLY LOW AMOUNT OF INSULIN AND SHOULD NOT HAVE FILLED THE INFUSION SET TUBING. DURING TROUBLESHOOTING THE PATIENT LOADED THE CARTRIDGE AND PRIMED 2.7 UNITS. THE PUMP EMITTED A "PUMP IS NOT PRIMED" WARNING AND HE ALSO PRIMED 0.6 AND 3 UNITS. THE LOW PRIME VOLUME COULD BE DUE TO THE PUMP DISPENSING INSULIN INTO THE INFUSION SET TUBING DURING THE LOAD STEP. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 46 YR