ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2011-05340
- Event Type
- Malfunction
- Date Received
- July 27, 2011
- Date of Event
- June 26, 2011
- Report Date
- June 26, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP WAS RETURNED TO ANIMAS. REVIEW OF THE PUMP HISTORY VERIFIED LOSS OF PRIME WARNINGS DURING THE LOAD STEP. THE PUMP WAS REWOUND, LOADED, AND PRIMED WITH NO ALARMS DURING INVESTIGATION. THE PUMP WAS EXERCISED WITH NO ALARMS OCCURRING. THE FORCE SENSOR CALIBRATION READING WAS OUT OF SPECIFICATION HOWEVER THE RESISTANCE READING WAS WITHIN SPECIFICATIONS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PATIENT CALLED TO REPORT THAT HE CHANGED HIS INFUSION SET AND WHEN HE PRIMED HIS PUMP HE ONLY PRIMED 9 UNITS. HE PREVIOUSLY PRIMED 21.6 UNITS ON (B)(6) AND 21.8 UNITS ON (B)(6) BUT ON (B)(6) HE ONLY PRIMED 9 UNITS TO SEE 5 DROPS OF INSULIN COMING OUT OF THE END OF THE TUBING. THE 9 UNITS IS AN UNEXPECTEDLY LOW AMOUNT OF INSULIN AND SHOULD NOT HAVE FILLED THE INFUSION SET TUBING. DURING TROUBLESHOOTING THE PATIENT LOADED THE CARTRIDGE AND PRIMED 2.7 UNITS. THE PUMP EMITTED A "PUMP IS NOT PRIMED" WARNING AND HE ALSO PRIMED 0.6 AND 3 UNITS. THE LOW PRIME VOLUME COULD BE DUE TO THE PUMP DISPENSING INSULIN INTO THE INFUSION SET TUBING DURING THE LOAD STEP. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR |