FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 SYNCHRON® SYSTEM

MDR report key: 2180477 · Received July 27, 2011

Report

Report Number
2050012-2011-03831
Event Type
Malfunction
Date Received
July 27, 2011
Date of Event
June 26, 2011
Report Date
June 26, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2011 FOR THIS EVENT. THE FSE VERIFIED DRIED SOLUTION ON AND BELOW THE REGULATOR. THE FSE REPLACED AND ADJUSTED THE 10 PSI AIR REGULATOR, DRAINED THE MIST SEPARATORS ON THE HYDRO AND PRIMED THE SYSTEM. NO FURTHER HYDRO LEAKING WAS REPORTED AS OF (B)(4) 2011. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED TO BECKMAN COULTER, INC. (BEC) A LEAK FROM THEIR UNICEL DXC 800 SYNCHRON SYSTEM. THE CUSTOMER REPORTED A RECURRING FLUID LEAKING FROM THE 10 PSI AIR REGULATOR. THE LEAK WAS FROM THE HYDRO AREA ONTO THE FLOOR. NO FUMES WERE PRODUCED AND THE CUSTOMER WAS NEITHER HARMED NOR NEEDED MEDICAL TREATMENT. BEC CUSTOMER TECHNICAL SERVICE (CTS) PERSONNEL ADVISED THAT LAB PERSONNEL AVOID CONTACT WITH THE FLUID AND TO NOT USE THE SYSTEM UNTIL THE SYSTEM IS EVALUATED BY BEC. NO EFFECT TO PATIENT OR USER WAS REPORTED IN CONNECTION WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 SYNCHRON® SYSTEM DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER FOR CLINICAL USE JJE BECKMAN COULTER, INC. DXC 800 PRO NA

Patients

Seq Age Sex Outcome Treatment
1