FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 2180430 · Received July 27, 2011

Report

Report Number
2050012-2011-03089
Event Type
Malfunction
Date Received
July 27, 2011
Date of Event
June 25, 2011
Report Date
June 25, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BECKMAN COULTER, INC. (BEC) TO REPORT ERRONEOUSLY HIGH MAGNESIUM (MG) RESULTS FOR 25 PATIENT SAMPLES ASSAYED ON THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. THE PATIENT RESULTS WERE REPORTED OUT OF THE LABORATORY. THERE WAS NO REPORT OF PATIENT TREATMENT BEING IMPACTED BY THE ERRONEOUSLY HIGH MG RESULTS. THE CUSTOMER REPORTED THAT QUALITY CONTROL RESULTS FOR MG WERE WITHIN THEIR ESTABLISHED RANGES PRIOR TO THE EVENT. THE CUSTOMER REPORTED THAT AN ERROR MESSAGE OF 10 CUVETTE OUT-OF-LIMITS AND AN OBSERVED LEAK FROM THE REAGENT PROBES PROMPTED THE CUSTOMER TO REVIEW THE ANALYZER RESULTS. THE MG RESULTS WERE QUESTIONED BECAUSE, THEY WERE ABOVE THE MG REFERENCE RANGE FOR THE LABORATORY. A FIELD SERVICE ENGINEER WAS DISPATCHED TO THE CUSTOMER'S SITE. THE FSE DISCOVERED A FAULTY FLOAT SWITCH IN THE WASH SOLUTION CANISTER. THE CANISTER HAD RUN DRY AND WAS NEVER REFILLED WHILE THE ANALYZER HAD CONTINUED TO OPERATE. THIS CAUSED THE CUVETTE OUT-OF-LIMIT ERRORS AND ALSO ALLOWED AIR INTO THE REAGENT LINES WHICH CAUSED THE REMAINING LIQUID IN THE PROBES TO LEAK OUT. THE FSE REPLACED THE FLOAT SWITCH WHICH CORRECTED THE PROBLEM. A 20 POINT PRECISION RUN FOR MG WAS PERFORMED WITH RESULTS WITHIN THE PRECISION CLAIM FOR THE MG ASSAY. QUALITY CONTROL RESULTS FOR MG WERE WITHIN THE LABORATORY'S ESTABLISHED RANGES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER JJE BECKMAN COULTER, INC. NA

Patients

Seq Age Sex Outcome Treatment
1 MAGNESIUM REAGENT