FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2180401 · Received July 27, 2011

Report

Report Number
6000001-2011-14851
Event Type
Malfunction
Date Received
July 27, 2011
Date of Event
June 30, 2011
Report Date
July 7, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH PUMP FAILURE WAS NOT CONFIRMED OR REPRODUCED DURING THE PRODUCT EVALUATION. HOWEVER, SERVICE FOUND FC 814:04 TO BE THE LAST ISSUE THE PUMP EXPERIENCED BEFORE SERVICE, AND WAS REPRODUCED UPON START UP. THE ROOT CAUSE OF THIS CONDITION WAS ASSIGNED TO A LOOSE CONNECTION BETWEEN THE PUMP HOUSE MODULE PRINTED CIRCUIT BOARD AND THE AIR IN LINE PRINTED CIRCUIT BOARD. THE AIR IN LINE PRINTED CIRCUIT BOARD WAS REPLUGGED TO CORRECT THIS CONDITION THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 5.09.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. A DEVICE HISTORY REVIEW WAS PERFORMED WITH NO EXCEPTIONS DURING MANUFACTURING FOUND.

Additional Manufacturer Narrative · 1

(B)(4).THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED A TECHNICAL SERVICE REPRESENTATIVE TO REPORT A COLLEAGUE INFUSION PUMP WITH PUMP FAILURE. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THIS CONDITION WAS FOUND IN THE BIOMED DEPARTMENT. IT IS UNKNOWN WHEN THIS EVENT OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT, PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION IS UNKNOWN AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1