LIGAMAX-5MM
Report
- Report Number
- 3005075853-2011-03025
- Event Type
- Malfunction
- Date Received
- July 27, 2011
- Date of Event
- June 27, 2011
- Report Date
- July 8, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING SPECIFICATIONS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. ADDITIONAL INFORMATION WAS REQUESTED AND THE FOLLOWING WAS OBTAINED: WHICH FIRING OF THE DEVICE DID THIS EVENT OCCUR ON? IT OCCURRED ON THE FIRST TIME THE SURGEON TRIED TO FIRE THE DEVICE. WHAT VESSEL OR STRUCTURE WAS THE DEVICE FIRED ON AT THE TIME OF THE EVENT? IT WAS MEANT TO SEAL THE DUCTUS CHOLEDOCHUS DURING A LAPAROSCOPIC CHOLECYSTECTOMY. WAS THE CLIP FULLY ADVANCED INTO THE JAWS PRIOR TO FIRING? YES, THE CLIP WAS FULLY ADVANCED BEFORE FIRING. WAS THERE ANY TORQUING OR TWISTING OF THE DEVICE PRESENT AT THE TIME OF FIRING? NO. WAS ANY UNEXPECTED RESISTANCE FELT WHILE FIRING THE TRIGGER? NO. WERE ANY UNEXPECTED NOISES HEARD? IF SO, WHEN? NO. DID ANYTHING UNEXPECTED HAPPEN PRIOR TO THIS INCIDENT? NO. WAS THE DEVICE FIRED AFTER THIS INCIDENT IN OR OUT OF THE PATIENT? NO.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE SURGEON TRIED TO FIRE THE CLIP BUT WAS NOT ABLE TO CLOSE THE CLIP COMPLETELY, THE CLIP WAS ONLY CLOSING ON THE TIP. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGAMAX-5MM | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |