FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2180373 · Received July 27, 2011

Report

Report Number
1423500-2011-09759
Event Type
Malfunction
Date Received
July 27, 2011
Date of Event
July 3, 2011
Report Date
July 4, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR AIR IN TUBING (WITHOUT ALARM) WAS NOT CONFIRMED DUE TO UNAVAILABLE SAMPLE. THE CAUSE WAS THAT THE PATIENT CONNECTED DURING THE PRIMING OF THE DISPOSABLE SET. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. RENAL QUALITY ENGINEERING, ALONG WITH PLANT QUALITY AND MANUFACTURING PERSONNEL, WILL CONTINUE TO MONITOR THIS PRODUCT LINE FOR TRENDS AND WILL TAKE CORRECTIVE/PREVENTIVE ACTION AS APPROPRIATE.

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE AVAILABILITY IS UNKNOWN.THE LOT NUMBER IS UNKNOWN. SINCE THE LOT NUMBER IS UNKNOWN, NO BATCH REVIEW WILL BE PERFORMED. A FOLLOW-UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS OBTAINED

Description of Event or Problem · 1

THIS REPORT IS TO ADDRESS THE ISSUE OF AIR IN PATIENT LINE. THE CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER TO REPORT A CHECK PATIENT LINE OCCURRED ON HOME CHOICE (HC) DURING USE DURING DRAIN 1 . THE BAXTER TECHNICAL REPRESENTATIVE (TSR) HAD THE HOME PATIENT (HP) CHECK LINE FOR KINKS AND OCCLUSION. THE HP STATED THAT THE LINE WAS FULL OF AIR BUBBLES. INITIAL DRAIN VOLUME (IDV)=3ML. THE TSR ASSISTED THE HP IN ENDING THERAPY AND ADVISED TO START OVER WITH NEW SUPPLIES. THERE WAS PATIENT INVOLVEMENT. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 62 YR HOMECHOICE